Portable COD Meter

Portable COD Meter
Details:
Model: QW-COD-B
Ambient Temperature: 5 - 40℃
Relative Humidity: < 85% (no condensation)
Power Supply: AC 220V±10% / 50Hz
Send Inquiry
Description
Send Inquiry

Hangzhou Qiwei Instrument Co., Ltd. is one of the leading manufacturers and suppliers of portable cod meter in China. With a professional production team, we are able to meet the needs of the majority of our customers. Welcome to wholesale bulk portable cod meter in stock here from our factory. We also accept customized orders.

 

 
 

 

Environmental water quality inspection often involves multiple sampling points rather than a single fixed laboratory location. Inspectors may need to collect wastewater samples from industrial discharge outlets, drainage channels, rivers and wastewater treatment facilities during the same inspection task.

The QW-COD-B Portable COD Meter is intended for COD measurement of collected water samples. It can be included in a mobile water testing arrangement for field sampling programs, wastewater discharge checks and preliminary water pollution investigations.

For multi-point inspection, operators can label each sample by location and collection time, prepare the required COD reaction system, complete digestion and cooling, and measure each sample using the analyzer. Keeping sample identification and preparation conditions consistent helps make results from different sampling locations easier to review.

This workflow is particularly useful for environmental service providers and inspection teams that need to organize COD testing across several monitoring locations.

 

product key technologies
product-800-600
01.

The portable COD analyzer is used to measure Chemical Oxygen Demand (COD). This indicator refers to the equivalent consumption of strong oxidizing agents needed to oxidize reducible substances in water samples under standardized test conditions. It is a vital index for quantifying reducible matter present in water. These reducible components mainly consist of organic compounds, together with nitrites, sulfides and ferrous salts. Therefore, COD acts as a major reference for evaluating organic pollution in water bodies. A higher COD value indicates a more severe level of organic contamination in the water sample.

02.

This portable meter comes with its own set of reagents, including a custom composite catalyst. The catalyst accelerates the oxidation reaction and helps reduce interference from chloride ions. Inside the digestion unit, the sample and reagents react rapidly through a redox process, producing chromium ions(Cr3+). The instrument then uses spectrophotometry to measure the concentration of these ions and calculate the COD level.

The entire system has been upgraded for better stability, accuracy, measuring range, and ease of use. It is designed to handle a variety of water testing tasks in different environments. The meter is suitable for analyzing wastewater from chemical plants, petrochemical facilities, coking operations, paper mills, metallurgical plants, breweries, and pharmaceutical factories. It also performs well for municipal sewage, household wastewater, and natural surface water from rivers and lakes.

product-600-600

Products Description

 

Uniform test workflows are critical when running COD measurements across multiple field sites. The QW-COD-B adopts rapid catalytic digestion combined with spectrophotometric detection. Operators must follow the standard reagent protocol and keep reaction conditions stable to guarantee comparable test outcomes.

Step 1: Collect Representative Water Samples

Gather water specimens from predefined monitoring points, including river sections, drainage ditches, or different stages inside wastewater treatment processes. Mark every sample clearly with its name, sampling timestamp and unique test ID.

Step 2: Set up the COD Reaction System

Prepare water specimens and matching COD reagents strictly following the test protocol and target measuring range. Make sure to use correct reagent dosages and compatible reaction vials. Pre-loaded digestion tubes tailored for the QW-COD-B simplify reagent preparation and sample handling during field operations.

Step 3: Carry out Digestion under Controlled Conditions

Load the prepared reaction vials into the digestion module. As specified in the QW-COD-B product specification, digestion runs for 20 minutes at a constant temperature of 165±1°C. Please comply with the official guidance for heating, cooling and sample tube placement.

Step 4: Spectrophotometric Detection

Once digestion finishes and the mixture cools down to ambient temperature, get the reacted solution ready for testing as required by the instrument operation guide. The analyzer reads the absorbance of reaction products via spectrophotometry and calculates the COD concentration.

Step 5: Log and Analyze Test Data

Record the COD reading together with sampling site, collection time, selected measurement range and relevant test parameters. For pollution surveys or discharge compliance checks, cross-reference these results against regulatory standards and other supporting water quality indicators.

 

Performance Indicator

Model

QW-COD-B

Ambient Temperature

5 - 40℃

Relative Humidity

< 85% (no condensation)

Power Supply

AC 220V±10% / 50Hz

Technical Parameter

Model

QW-COD-B

Method Of Determination

Rapid catalytic method (chromium method)

Measuring Range

0~150 mg/L low range; 100~1000 mg/L high range 

(diluted measurement is required when the concentration is above 1000 mg/L)

Measuring Error

5~100 mg/L, absolute error ≤ ±5 mg/L; 100~1000 mg/L, relative error ≤ ±5%

Dissolution Temperature

165 ± 1℃

Dissolution Time

20 minutes

DeterminationTime

20 minutes. Can be used to measure 9 wells or 25 wells simultaneously 

(depending on the configuration of the digestion instrument)

Temperature Indication Error

<±1℃

Temperature Field Uniformity

≤ 2℃

Indication Error

≤ ±2%

Record Storage

It can store 20 measurement results

How to Organize a Mobile COD Testing Task

 

A field COD testing task requires more than the analyzer itself. Before leaving the laboratory or testing station, operators should prepare the required equipment, consumables and sample identification materials.

Before Field Deployment

Prepare the QW-COD-B, compatible COD digestion equipment, reaction tubes, required reagents, sample containers and suitable power supply. Confirm that the selected digestion equipment is compatible with the analyzer and the intended test method.

At the Sampling Location

Identify each sampling point and record the collection time. Collect representative water samples using the applicable sampling procedure. For pollution investigation, keep samples from different locations separately identified to avoid mixing up their origins.

During COD Testing

Prepare the reaction system according to the selected measurement range and reagent procedure. Complete digestion, cooling and spectrophotometric measurement as instructed. Keep the same preparation conditions when comparing samples collected from different locations.

After Testing

Record the COD result, sample identification, sampling location and test time. Clean the relevant accessories and organize used reaction tubes and chemical waste according to applicable laboratory and environmental safety procedures.

A well-organized field testing procedure helps inspection teams reduce sample identification errors and maintain consistent testing records.

 

Application

 

QW-COD-B is a field COD analyzer built for on-site chemical oxygen demand testing. It serves a wide array of application scenarios such as environmental audits, pollution investigations and routine wastewater monitoring, and is ideal for organizations carrying out multi-site sampling and fast field analysis close to sampling points.

1. Environmental Regulation and Law Enforcement

Environmental inspection teams may use this portable COD device to analyze water specimens collected from industrial discharge points, drainage trenches, municipal sewage facilities and suspected pollution sites. Field tests help inspectors obtain raw on-site data, which provides solid support for case tracing and follow-up sampling plans.

2. Emergency Water Pollution Handling

When a suspected water contamination incident takes place, monitoring staff often need to collect samples from upstream reaches, downstream areas and discharge outlets within a limited timeframe. The QW-COD-B can be packed as part of a mobile testing set to get COD results quickly once samples are collected and preprocessed. For emergency assignments, prepare power sources, digestion modules, reagent kits, reaction vials and sample-handling accessories in advance. Test data can be cross-referenced alongside other water quality parameters. If official confirmation is needed, standard sampling protocols and laboratory verification can be implemented afterwards.

3. On-site Inspection at Wastewater Treatment Plants

Workers at treatment facilities can collect samples from the inlet, each treatment stage and final discharge end to track COD changes during the purification process. Rapid field checks make it easier to catch unusual value shifts early and offer extra reference information for process diagnosis.

4. Monitoring of Industrial Wastewater Discharge

Manufacturers from machinery, pulp & paper, petrochemical, fermentation and pharmaceutical sectors need regular wastewater monitoring at different production and discharge nodes. This analyzer is capable of measuring COD for both dilute and highly concentrated effluent samples.

5. Field Sampling of Rivers, Lakes and Surface Water

Environmental regulators and water management teams can include this COD analyzer in their mobile sampling toolkit for rivers, lakes and other natural open water. On-site test outcomes help assess organic pollution levels at randomly selected monitoring locations.

6. Third-party Environmental Testing Services

Commercial environmental laboratories and professional testing service providers can assign this portable COD instrument to their field sampling teams. It enables preliminary screening and quick on-site analysis before samples are sent back to the lab for formal, comprehensive testing.

 