Lab High Power Digital High Shear Emulsifier Timed

Lab High Power Digital High Shear Emulsifier Timed
Details:
Model: JRH-400S
Power: 400W
Time Setting: Timed
Stator Size: φ65mm
Send Inquiry
Description
Send Inquiry

Hangzhou Qiwei Instrument Co., Ltd. is one of the leading manufacturers and suppliers of lab high power digital high shear emulsifier timed in China. With a professional production team, we are able to meet the needs of the majority of our customers. Welcome to wholesale bulk lab high power digital high shear emulsifier timed in stock here from our factory. We also accept customized orders.

 

 
 
product key technologies
product-800-800

The Lab High Power Digital High Shear Emulsifier Timed adopts custom-designed rotor and stator components. Under high-speed motor operation, materials to be processed are sucked into the rotor chamber. The high-frequency mechanical motion generated by the rotor's high-speed rotation delivers tremendous kinetic energy. In the tiny precise clearance between stator and rotor, materials experience intensive mechanical and fluid shearing, centrifugal squeezing, interfacial friction, high-speed impact rupture and turbulence, realizing material splitting, grinding and dispersion. Hundreds of thousands of repeated shearing acts on materials in an instant, enabling immiscible materials to attain homogeneous and fine emulsification, crushing and dissolving effects. The Lab High Power Digital High Shear Emulsifier Timed is widely used in adhesives, architectural coatings, cosmetics, food processing, medicine, plastic resins, textile printing and dyeing, printing inks, asphalt and other industrial sectors.

 

Technical Parameter

 

Model

JRH-400S

JRH-1000S

JRH-400

JRH-1000

Power

400W

1100W

400W

1100W

Time Setting

Timed

Stator Size

φ65mm

Speed Adjustable Range

100-8000rpm

Range of Vertical Stroke

270mm

Power Supply

220V 50HZ

Length of the Stirring Rod

550mm

Size

460×420×750mm

Weight

38KG

Lifting Method

Automatic lifting and lowering

Hand-operated lifting and lowering

 

FAQ

Q: Is there any rule about running the machine without materials?

A: Never start the device when it is idling. Make sure the feed inlet at the bottom of the working head is fully submerged in the test liquid before powering it on. Running the unit with no liquid inside will cause severe friction during high-speed operation, which will burn out the bearings and mechanical seals.

Q: What should we pay attention to when using the timer function?

A: High-speed shearing for a long time will generate plenty of heat. If your samples are sensitive to temperature, use an ice water bath or other cooling methods. Do not let the machine run continuously for too long, so as to prevent the motor from overheating and the materials from going bad.

Q: What requirements do we have for liquid level during operation?

A: Keep the working head at a proper distance from the bottom and side walls of the container. The liquid level must stay above the stator holes of the working head. Improper liquid level will lead to liquid splashing or insufficient material feeding.

 

 

 

 