Ion-Selective Electrode

Ion-Selective Electrode
Details:
Electrode: Fluoride, Chloride, Bromide
Model: F201, CL201, Br201
Description: Crystal, Solid state, Solid state
Optimum PH Range: 5~6, 2~12, 2~11
Send Inquiry
Description
Send Inquiry

Hangzhou Qiwei Instrument Co., Ltd. is one of the leading manufacturers and suppliers of ion-selective electrode in China. With a professional production team, we are able to meet the needs of the majority of our customers. Welcome to wholesale bulk ion-selective electrode in stock here from our factory. We also accept customized orders.

 

 
 
product key technologies
Ion-Selective Electrode2

The Ion-Selective Electrode is a high-precision and high-stability professional sensing element. It is specially designed for the rapid and accurate determination of specific ions in aqueous solutions. Widely applicable to laboratory analysis, industrial online monitoring, field on-site detection and other scenarios, it serves as an indispensable core testing accessory for scientific research and teaching, production quality control, environmental monitoring and other fields.

 

Technical parameters

 

Electrode

Fluoride

Chloride

Bromide

Model

F201

CL201

Br201

Description

Crystal

Solid state

Solid state

Measuring Range

1~5×10-7MF(19000~0.01ppm)

10-2~5×10-5MCl (35500~1.8ppm)

1~5×10-6MBr(79000~0.4ppm)

Optimum PH Range

5~6

2~12

2~11

Responds Time

<5min

<2min

<2min

Membrane Resistance

<10MΩ

<0.5MΩ

<0.5MΩ

Temperature Range

5~60℃

0~60℃

0~60℃

Potentiometric Selective Coefficients

OH-

I-(2×102); NO3-(10-4); SO42-(10-4); Br-(4×102); CN-(2×102); Grave interference by sulphurion

I-(2.6); Cl-(1×10-5); NO3-(3.8×10-4); SO42-(10-4); Grave interference by sulphurion

Standard Solution

NaF

NaCl

NaBr

 

Electrode

Iodide

Cyanide

Mercury

Model

I201

H201

Hg201

Description

Solid state

Solid state

Solid state

Measuring Range

10-2×5×10-7MI(127000~0.06ppm)

10-2~10-6MCN(260~0.03ppm)

10-1~5×10-6MHg2+ (20000~1ppm)

Optimum PH Range

2~12

≥7

3~5

Responds Time

<2min

<2min

<2min

Membrane Resistance

<0.5MΩ

<0.5MΩ

<0.5MΩ

Temperature Range

0~60℃

0~60℃

0~60℃

Potentiometric Selective Coefficients

Cl-(2×10-6); Br-(1×10-6); NO3-(2×10-7); SO42-(7×10-7); HPO42-(8.3×10-7)

Main interference ions: S2-; I-; Hg2+

Interference by silver ions

Standard Solution

KI

NaCN

Hg2+

 

Electrode

Silver/Sulfide

Plumbum

Cuprum

Model

Ag201

Pb201

Cu201

Description

Solid state

Solid state

Solid state

Measuring Range

10-1~10-7M
(108000~0.01ppm) Ag+
(3200~0.003ppm) S2-

10-1~5×10-7MPb2+ (20700~0.1ppm)

1~5×10-7MCu2+ (63500~0.012ppm)

Optimum PH Range

2~12
>12

3~6

3~5

Responds Time

<2min

<2min

<2min

Membrane Resistance

<0.5MΩ

<0.6MΩ

<0.5MΩ

Temperature Range

0~60℃

0~60℃

0~60℃

Potentiometric Selective Coefficients

Noninterference ions: F+; NO3-; I-; Cl-

Fe3+(2.7); Cu2+(1×103); Mg2+(4×10-6); Ba2+(2×10-5); Ca2+(6×10-6); Zn2+(2×10-5)

Noninterference ions: K+; Ca2+; Ba2+; Zn2+; Fe2+
Interference ions: S2-; Fe3+

Standard Solution

Silver: AgNO3
Sulfide: NaS

PbNO3

CuNO3

 

Electrode

Cadmium

Micro-Kalium

Calcium

Model

Cd201

K401

Ca201

Description

Solid state

PVC

PVC

Measuring Range

10-1~5×10-7MCd2+ (11200~0.05ppm)

10-1~5×10-7MK(3900~0.02ppm)

10-1~5×10-6MCa2+ (4000~0.2ppm)

Optimum PH Range

3~10

3.5~10.5

5~10

Responds Time

<2min

<3min

<3min

Membrane Resistance

<0.5MΩ

<30MΩ

<10MΩ

Temperature Range

0~60℃

0~45℃

0~45℃

Potentiometric Selective Coefficients

Noninterference ions: K+; NaInterference ions: Hg2+; Pb2+; Ag+; S2-

Na+(3×10-3); Ca2+(1×10-5); Li+(1×10-3)

K+((1×10-4); Na+(1×10-4); Zn2+(7×10-1); Ba2+(3×10-5); Mg2+(1×10-3); Fe3+(1)

Standard Solution

CdNO3

KCl

CaCl

 

Electrode

Nitrate

Fluoroborate

Vater hardness

Model

NO3201

FPSG2

YGa201

Description

PVC

PVC

PVC

Measuring Range

1×10-6MNO3(62000~0.062ppm)

10-1~3×10-6MBF4(8600~0.3ppm)

10-1~5×10-5M

Optimum PH Range

2.5~9

3~11
(10-1~10-4MBF4-)

2~9

Responds Time

<3min

<3min

<3min

Membrane Resistance

<10MΩ

<1MΩ

<10MΩ

Temperature Range

5~45℃

5~40℃

0~40℃

Potentiometric Selective Coefficients

Cl-(4.9×10-2); SO42-(4×10-5); H2PO3-(1×10-4)

SO42-(2.5×10-6); F-(3×10-6); Br-(10-3); NO3-(6.5×10-3); Cl-(6×10-5); I-(8×10-2); CNS-(0.25)

Na+(1); K+(1×10-1)

Standard Solution

NaNO3

NaBF4

Ca2+, Mg2+

 

Electrode

Ferric

Ammonia

Carbon dioxide

Model

Fe201

A201

CO2

Description

FeCl4 PVC

Gas sensor

Gas sensor

Measuring Range

10-1~1×10-4MFe3+ (35500~1.8ppm)

1×10-6M (18000~0.018ppm)

10-2~5×10-5M (480~1ppm)

Optimum PH Range

<2

>11

/

Responds Time

<2min

<5min

<5min

Membrane Resistance

<2MΩ

<300MΩ

<300MΩ

Temperature Range

0~45℃

0~60℃

0~60℃

Potentiometric Selective Coefficients

NO3-, SO42-, Cl-
Zn2+ < Hg2+ < Fe2+ < Sn2+

Inteference by volatile amines

Interference by volatile weak acid

Standard Solution

Fe3+

NH4Cl

NaHCO3

 

Electrode

Hydrothion

Sulfur dioxide

Sodium

Model

LHH2

RYHL2

Na201

Description

Gas sensor

Gas sensor

Glass

Measuring Range

10-2~5×10-7M (480~1ppm)

10-2~10-5M

1~10-7MNa+ (23000~0.002ppm)

Optimum PH Range

/

/

/

Responds Time

/

/

/

Membrane Resistance

/

/

<50MΩ

Temperature Range

/

/

5~80℃

Potentiometric Selective Coefficients

/

/

H+(20~30) K(1/20~1/30)

Standard Solution

/

/

 

Hot Tags: ion-selective electrode, China ion-selective electrode manufacturers, suppliers, factory, electrode, Industrial Online PH Electrode, Ion Selective Electrode, PH Corrosion Resistant Electrode, PH Flat Electrode, PH Glass Composite Electrode

Send Inquiry
Send Inquiry